MG125數顯絕緣電阻表|兆歐表@MG125%
低壓3量程式| MG125 | 測量量程 | 最高準確度 | 分辨率 |
| MΩ | 4M/40M/400M/4000M | ±(3%+4) | 0.01MΩ |
| 附加電壓 | 125/50/25V | ||
| ACV/DCV | 600V(AC/DC自動識別) | ±(3%+2) | 1V |
| Ω | 4000Ω(蜂鳴與報警聯動) | ±(3%+3) | 1Ω |
| Ω | 40Ω | ±(3%+10) | 0.01Ω |
| 無負荷電壓 | 額定電壓的1-1.3倍 | ||
| 額定測量電流 | 1.0-1.2mA | ||
| 短路電流 | 2mA以下 | ||
| 線路帶電狀態測量 | 30VAC/DC 以上的測試負荷,發出蜂鳴音且報警等亮燈 | ||
| 內置電池 | LR6(5號堿性)*6 | ||
| 尺寸/重量 | H170*W142*D57mm/約600g | ||
| 附件 | 測試線(TL-112),便攜式(ST-50),使用說明書 | ||
MG125 digital insulation resistance meter | megohm table @MG125%
Low voltage 3 program
Measuring insulation resistance when additional voltage: 3 range switching type DC125V/50V/25V
Based on the safety design of Japanese industrial standard C1302-2002
The insulation resistance measurement function can automatically identify the wire charged state
Use large font and analog bar graph is easy to read
After measurement and can be automatically locked the finalmeasurements
Liquid crystal display with backlight
With automatic discharge function
AC/DC600V voltage measurement with function of automatic recognition of AC/DC type
With the 4000/40 resistor measurement range of conducting buzzeralarm and linkage
Numerical display: 4000
Sample rate: about 2 times / sec
Safety standards: IEC61010 CAT III600V MG125 measurement range ofthe highest accuracy resolution
M Ω 4M/40M/400M/4000M ± (3%+4) 0.01M.
Additional voltage 125/50/25V
ACV/DCV 600V (AC/DC automatic identification) ± (3%+2) 1V
Omega 4000 omega (beep and alarm linkage) + (3%+3) 1.
Omega 40 Omega + (3%+10) 0.01.
No load voltage of 1-1.3 times rated voltage
Rated 1.0-1.2mA current measurement
Short circuit current below 2mA
Live line measurement 30VAC/DC the above test load, a beep soundand light alarm etc.
The built-in battery LR6 (5 alkaline) *6
Size / weight about H170*W142*D57mm/ 600g
Accessories test line (TL-112), portable (ST-50), instructions for use