Inspector手持式表面污染檢測(cè)儀
簡介Inspector hand-held surface contamination detector
Introduction
· Inspector detector can be used to detect alpha, beta,gamma and X rays
· sensitivity built-in 2 inch flat GM detector to alpha andbeta ray source is very high
Digital display can be selected according to the requirements of mR/hr, CPM, CPS or m Sv/hr differentunits
· detection dead time automatic compensation of GM pipe, anti saturation circuit to prevent interference withstrong radiation field
· tester in accordance with the European CE certificationrequired
· the detector can detect the slight pollution, can be timedreadings
Time can be set from 1 to 24 hours, the calibration to the "safety first" as the standard to avoid damage checkingpersonnel radiation leakage
Technical specifications
· detector halogen quencher GM tube, effective diameter 45mm, mica window density 1.5-2.0 mg/cm3
· display four digit display by LCD screen, count light redLED lamp
· the average cycle are updated once per 3S, shows the average value of 30s. If the radiation intensity, the average 30s period will be shortened. The user can choose the shortest average cycle 3S
Operating range of 0.001 to 100 mR/hr, 0 to 300000 CPM, count a total of 1 to 9999000; 0.01 to 1000 μ Sv/hr, 0 to 5000 CP
· efficiency of Sr-90 (546KeV, 2.3MeV β max) efficiency of about 75%; C-14 (156KeV β max) efficiency of about 11
Bi-210 (1.2 MeV β max) efficiency of about 64%; Am-241(5.5 MeV a) efficiency of about 36
· timer can be set up to receive radiation time of 1 to 10min, increments 1min; or 10-50min, increment 10min;or 1-24hr, increment 1hr
· power source 1 9V alkaline batteries, the battery can be used in natural under 200H, in the 1 mR/hr dose rate at least available 24h
· output double micro jack driven CMOS or TTL program,data can be downloaded to a computer. Micro jack allows for electronic check
Anti saturation of the reader in the field can display up to100 times the maximum readings to full scale
· voice the voice prompts, you can set to mute
· volume and weight of 150 × 80 × 30 mm, 350 g with battery
· sensitivity is 3500 CPM/mR/hr, for the 137Cs standardsource
· accuracy ± 15%
· the ambient temperature from -10 to +50oC
· options can wipe test plate, external scintillator
Inspector type EXP detector
Inspector EXP tester is designed for individual emergency use
· and Inspector detector has the same performance, small size, detector of good quality; but it is the use of an external probe
· there are external probe have some functional properties
· carrying case with a transparent window, through which you can see the whole counter reading
And at the same time it can also adjust the control switch
· instrument box also includes a convenient cover, belt loop and a small bag of calibration source
If needed, the detector can be attached by single hand incarrying case
· detector protective cover can be conveniently folded,convenient operation
· detector and Inspector
· anodized aluminum and is provided with a black coated foam handle
· 500V power sources, a detector head
· volume 22.2 cm long, 7 cm in diameter
· is connected with the Amphenol 31226 twinax